Technical data

N7280
N7280N7280
N7280-
--
-90034
9003490034
90034-
--
-
Rev. A
Rev. ARev. A
Rev. A
System
SystemSystem
System Installation and Verification
Installation and Verification Installation and Verification
Installation and Verification
25
2525
25
Figure 21: Vacion power supply
4
44
4 Check interlocks — Enable the X-ray key and press the
ON switch. See Figure 22. Tap each of the 6 side access
panels, 2 topside access panels, tower door and interlock
points with your hand. Watch to see if the red interlock
LED on the control panel goes off. If it does, you will
need to adjust interlock switches.
For troubleshooting information, see Service
Documentation in the Medalist x6000 help system or
call the Agilent Support Center.
Vacion pump
Power Supply
Input
~200-240 V 60VA
Power
On
J04000
Measure the voltage
here (J04000)