Technical data

N7280
N7280N7280
N7280-
--
-90034
9003490034
90034-
--
-
Rev. A
Rev. ARev. A
Rev. A
System
SystemSystem
System Installation and Verification
Installation and Verification Installation and Verification
Installation and Verification
24
2424
24
c
cc
c Power on the system controller. Allow the system
controller to boot up and note any warnings or
errors. Errors indicate a system
(hardware/software) failure and require
troubleshooting and repair before proceeding. For
troubleshooting information, see Service
Documentation in the Medalist x6000 help system
or call the Agilent Support Center.
d
dd
d Set the system clock to the customer’s time zone.
e
ee
e Switch the PDU rocker switch on.
3
33
3 Vacion™ power supply voltage check — Use a
voltmeter, set on DC Millivolt, to measure the Vacion
power supply at test point J04000 on the power supply.
It is located under the main unit in the rear. See Figure
21. Measure this at the BNC connector on the power
supply. The reading may be as high as 3 volts d.c. but
after warm-up (~1/2 hour) it should read in the 20 - 50
millivolt range and be continuing to drop slowly.
Do not proceed with startup of the Medalist x6000 until this reading is below 20 mv. Contact
the Agilent Customer Support Center for help is the reading remains above 20mv.
CAUTION