Technical data

N7280
N7280N7280
N7280-
--
-90034
9003490034
90034-
--
-
Rev. A
Rev. ARev. A
Rev. A
System
SystemSystem
System Installation and Verification
Installation and Verification Installation and Verification
Installation and Verification
2
22
2
2
2 2
2 Preparing for Agilent Medalist x6000 Installation
Preparing for Agilent Medalist x6000 InstallationPreparing for Agilent Medalist x6000 Installation
Preparing for Agilent Medalist x6000 Installation
You will need to refer to the following documents.
Unpacking Instructions (N7280-90010)
Site Preparation Manual (N7280-90009)
Agilent Medalist x6000 Help
Agilent Electrostatic Discharge Control Specification (A-
5951-1589-1)
You will need a copy of the x6000 X-ray Safety Test
Procedure (N7280-90093) and the x6000 X-Ray Safety
Test Report (N7280-90076) to complete the installation.
You need to acquire these before entering the customer
site. These documents are available at
http://www.agilent.com/find/x6000. Click Technical
Support and then click Download X-Ray Safety Test
Procedure.
The unpacking instructions (N7280-90010) will be
located on the outside of the crate. The Site Preparation
Manual has been sent to the customer before this
installation began. The Agilent Medalist x6000 help is
on-line when you get the system up and running. The
ESD document is one of general corporate nature that
you should be familiar with already. If you are not,
please review it as soon as possible.