Equipment that appears damaged or defective should be
made inoperative and secured against unintended operation
until they can be repaired by qualified service personnel.
See the Lock Out /Tag Out procedure in the Service section
of the on-line help.
Do Not Service or Adjust Alo
Do Not Service or Adjust AloDo Not Service or Adjust Alo
Do Not Service or Adjust Alone
nene
ne
Do not attempt internal service or adjustment of this
instrument unless another person, capable of rendering
first aid and resuscitation, is present.
D
DD
Do Not Subs
o Not Subso Not Subs
o Not Subst
tt
titute Parts or Modify Equipme
itute Parts or Modify Equipmeitute Parts or Modify Equipme
itute Parts or Modify Equipment
ntnt
nt
Because of the danger of introducing additional hazards, do
not install substitute parts or perform any unauthorized
modification to the product. Contact your Agilent
Authorized Service Representative for service and repair to
ensure that safety features are maintained.
Maintain the System as Recommended
Maintain the System as RecommendedMaintain the System as Recommended
Maintain the System as Recommended
The recommended preventive maintenance for the system is
documented in the on-line help. Preventive maintenance
should always include testing the emergency shut down
(EMO) switch by pressing it and verifying that all ac and dc
power turns off.
Fol
FolFol
Foll
ll
low All X
ow All Xow All X
ow All X-
--
-ray Safety Requirements and Legislation
ray Safety Requirements and Legislationray Safety Requirements and Legislation
ray Safety Requirements and Legislation
DO NOT operate the equipment unless an X-ray safety
survey has been performed within the last six (6) months,
when the Medalist x6000 is installed at a site, when the
Medalist x6000 has been moved, and immediately following
maintenance that involves the safety interlock subsystem,
the lead shielding or the X-ray source.
Do not operate this equipment without complying to these
recommendations.
Owners/Operators of this equipment are responsible for
compliance with all local legislation regarding X-ray
equipment.
An Operator’s Guide and Declaration of Conformity in a
local language is available upon request to Agilent
Technologies, Inc.