Technical data

34410A/11A/L4411A Service Guide 47
Quick Start 2
To Make a 2-Wire Temperature Measurement
Press to select the temperature function. Then press
and select RTD-2W or THERMISTOR-2W from the menu.
Probe types: 2.2 kW, 5 kW, 10 kW thermistors; 0.00385%/ºC RTD
Configurable parameters: PROBE TYPE, THERMISTOR or RTD value, AUTO ZERO,
OFFSET COMP (RTD probes only), INTEGRATION, UNITS, NULL, and NULL VALUE
To Make a 4-Wire Temperature Measurement
Press to select the temperature function. Then press
and select RTD-4W or THERMISTOR-4W from the menu.
Probe types: 2.2 kW, 5 kW, 10 kW thermistors; 0.00385%/ºC RTD
Configurable parameters: PROBE TYPE, THERMISTOR or RTD value, OFFSET COMP
(RTD probes only), INTEGRATION, UNITS, NULL, and NULL VALUE
All 4- wire temperature measurements are made with auto- zero on.
Connect test leads as shown:
Thermistor or RTD
Connect test leads as shown:
Thermistor or RTD