Technical data

This is a standard SCPI command. Please refer to the SCPI specification for details.
*RCL
IAgilentN490xSystem.RecallState (IVI-compliant)
*RCL <NR1 | string>
This command loads the setup from a numbered store or from a full path filename
that was previously stored with “*SAV ” on page 97. The range of store numbers
is 0 through 9.
In addition, upon receipt of *RCL, the error detector will reset all past results to
zero.
N O T E
Depending on the patterns that are saved with the setup, the instrument may require
up to half a minute to settle. See Allowing Serial BERT to Settle - Procedures” on
page 22 for details.
*SAV
IAgilentN490xSystem.SaveState (IVI-compliant)
*SAV <NR1 | string>
This command saves the current instrument setup into a numbered store or into
a full path filename. The range of store numbers is 0 through 9. The command
“*RCL ” on page 97 restores the setup.
The setup saves the currently used patterns, signal definitions, and other user
interface settings.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 97