Technical data
TEST:EXECute?
IIviDriverUtility.SelfTest (IVI-compliant)
TEST:EXECute? [SelfTest_value] {,<SelfTest_value>}
This command runs user-specified self tests. If no parameter is specified, all tests
are run.
Successful completion of a self test returns 0. If a self test fails, 1 is returned.
SelfTest_value can be one of the parameters listed below.
Table 104
Parameter Description
ALL Error detector and pulse generator
module self test is started.
PGENerator Pulse generator module self test is
started.
EDETector Error detector module self test is started.
PGCal Auto calibration of pulse generator delay.
EDCal Auto calibration of error detector delay.
N O T E
Use TEST:MESS? to read the result of the self tests.
TEST:MESSages?
IIviDriverUtility.ErrorQuery (IVI-compliant)
TEST:MESSages? PGPOn | EDPOn | EDET | PGEN
Returns a comma-separated list of messages. This command has the following
options:
• PGPOn: Pattern Generator Power On messages
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 403
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description