Technical data

EXTR, 9.910000000000e+037,
OMA,9.910000000000e+037,
AVGP,9.910000000000e+037
Parameters
The following SCPIs are both Commands and Queries, the input parameter is for
the command and the output parameter is for the queries:
:MEASure:EMASk(*):PARameter:TTIMe
:MEASure:EMASk(*):PARameter:TTIMe
:MEASure:EMASk(*):PARameter:TTIMe?
Sets the transition time.
Transition time, can be either 8020 or 9010.
Transition time, can be either 8020 or 9010.
:MEASure:EMASk(*):PARameter:WIDth
:MEASure:EMASk(*):PARameter:WIDth CROssing | CUStom
:MEASure:EMASk(*):PARameter:WIDth?
Defines the rule to calculate the eye width. It can be either at the eye crossing
point (CROssing) or custom defined (CUStom). Use “:MEASure:EMASk
(*):PARameter:WIDth:CUStom” on page 399 to define a custom percentage.
Example:
MEAS:EMAS2:PAR:WID CUS
MEAS:EMAS2:PAR:WID:CUS 40
These examples will calculate the eye width at 40% above the 0-Level.
Example:
meas:emas2:par:wid?
CRO
5 SCPI Command Reference
398 Agilent J-BERT N4903B High-Performance Serial BERT
Syntax
Description
Input
Output
Syntax
Description
Input