Technical data

:MEASure:EMASk(*):MARKer?
:MEASure:EMASk(*):MARKer?
Returns the threshold delay and the BER at the left and right intersections of the
marker lines.
x coordinate of left marker intersection, y coordinate of left marker intersection, x
coordinate of right marker intersection, y coordinate of right marker intersection.
Example:
meas:emas2:marker? 90,50,260,220
First four values are used for resetting the maker. The following six values are: Left
threshold value, Left time interval value, Left BER, Right threshold value, Right time
interval value, and Right BER.
Example:
47,243,89,265,-3.901291285419E-1,-4.157517291605E-1,4.999999999537E-1,3.021
65216 6406E-1,5.501573905349E-1,2.804230720563E-1
:MEASure:EMASk(*):MASK:LOAD?
:MEASure:EMASk(*):MASK:LOAD?
Loads the Mask File from the mask file folder provided with the software.
You need to run the measurement before loading a mask file. The installed mask
files are all valid.
Example:
meas:emas:cre?
2
meas:gen2:go
meas:emas2:mask:load? “c:01xGbEthernet.msk”
Returns the following values: MARGIN | NOMARGIN, Mask Alignment, Signal time
period is specified in the mask file , and waveform time period is 1/ED Clk Rate.
In the input file, if min_margin and max_margin are defined for all regions, then
the output is MARGIN; and using the margin SCPIs the mask test can resize the
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 385
Syntax
Description
Input
Output
Syntax
Description
Input