Technical data
+000,0.000000000000e+000,0.000000000000e+000,0.000000000000e+000…and
so on
:MEASure:EMASk(*):CLEAr
:MEASure:EMASk(*):CLEAr
This command clears the eye display data, and restarts the measurement.
None
None
:MEASure:EMASk(*):EXPort
:MEASure:EMASk(*):EXPort
This command exports the measurement data into the directory, and file name
specified by the user. The exported data is in the CSV format.
For exporting the data you need to fetch the data first.
There are five parameters:
• c:\test.txt : File name with complete Path
• ALL : You want to export only BER or ALL (BER, ComparedBits , Errored Bits)
• ALL : Want to consider ALL, expected 0s, expected 1s or any combination of
them
• 9 : Delimeter (9 for tab, 44 for comma , 59 for semicolon) (our exported data
can be comma, tab or semicolon separated, these are the corresponding ASCII
values)
• 0 : Clipboard option. 0 for No and 1 for Yes
Example:
meas:emas2:exp "c:\test.txt",ALL,ALL,9,0
None
5 SCPI Command Reference
384 Agilent J-BERT N4903B High-Performance Serial BERT