Technical data
This query returns the highest frequency of the frequency range of the selected
standard.
MEAS:JTOL:COMP(*):PAR:STANdard:NAMes?
IAgilentN490xJComplianceMeasurement.GetNames
MEAS:JTOL:COMP(*):PAR:STANdard:NAMes?
This query returns a list of all defined standards.
• Every standard provided by Agilent is identified by an abbreviation (<standard
enum>) and the standard's name (<name>).
For example:
<standard enum> :
CEI6GSR
<name> :
CEI 6 Gb/s Short Reach
• When a user-defined standard is added to the list of predefined standards, a
<standard enum> is automatically assigned.
The <standard enum> is used for the following commands
• MEAS:JTOL:COMP(*):PAR:STANdard[:SELect]
• MEAS:JTOL:COMP(*):PAR:STANdard:PLOT?
• MEAS:JTOL:COMP(*):PAR:STANdard:DESCription[?].
MEAS:JTOL:COMP(*):PAR:STANdard:PLOT?
IAgilentN490xJComplianceMeasurement.GetStandardData
MEAS:JTOL:COMP(*):PAR:STANdard:PLOT? [<standard enum>]
Returns a comma-separated list of jitter frequencies (Hz) and associated jitter
amplitudes (UI) of the standard specified by <standard enum>. If no input
parameter is given, the list of the currently selected standard is returned.
The list of <standard enum> is returned by the MEAS:JTOL:COMP
(*):PAR:STANdard:NAMes? query.
The <standard enum> of the currently selected standard is returned by means of
the MEAS:JTOL:COMP(*):PARameter:STANdard[:SELect]? query.
5 SCPI Command Reference
372 Agilent J-BERT N4903B High-Performance Serial BERT
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax