Technical data

MEAS:JTOL:COMP(*):PAR:STANdard:NAMes? query lists <standard enum>,
<name> pairs, where <name> is the standard's name.
The query returns the <standard enum> of the currently selected standard.
MEAS:JTOL:COMP(*):PAR:STANdard:USERdefined[?]
IAgilentN490xJComplianceMeasurement.UserDefinedStandardPoints()
MEAS:JTOL:COMP(*):PAR:STANdard:USERdefined (<NR3>, <NR3>, …)
MEAS:JTOL:COMP(*):PAR:STANdard:USERdefined?
The command specifies a user-defined standard. The input parameter is a comma-
separated list of jitter frequencies (Hz) and associated jitter amplitudes (UI). For
example, (2e3, 110, 3.24e4, 50, 71.234e4, 10, … ).
The <standard enum> of a standard loaded with this command is automatically
set to "USER"
Entering a name or a description for the standard is only possible via the user
interface. However, the description can be modified by means of the
MEAS:JTOL:COMP(*):PAR:STANdard:DESCription command.
The query returns a comma-separated list of the current user-defined standard.
MEAS:JTOL:COMP(*):PAR:STANdard:FMIN?
IAgilentN490xJComplianceMeasurement.GetFreqMin()
MEAS:JTOL:COMP(*):PAR:STANdard:FMIN?
This query returns the lowest frequency of the frequency range of the selected
standard.
MEAS:JTOL:COMP(*):PAR:STANdard:FMAX?
IAgilentN490xJComplianceMeasurement.GetFreqMax()
MEAS:JTOL:COMP(*):PAR:STANdard:FMAX?
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 371
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax