Technical data
MEAS:JTOL:COMP(*):PARameter:BTESt:NBITs[?]
IAgilentN490xJComplianceMeasurement.NumberOfBits
MEAS:JTOLerance:COMP(*):PARameter:BTESt:NBITs <NR3>
MEAS:JTOL:COMP(*):PAR:BTESt:NBITs?
This command is needed if the verification method is set to ABSolute. It sets the
number of bits to be captured and compared, for example 5e9. As soon as this
number is exceeded, the measurement proceeds to the next frequency. The setting
takes effect when the mode ABSolute is enabled by means of the
MEAS:JTOL:COMP(*):PAR:BTESt command.
The query returns the present number of bits.
MEAS:JTOL:COMP(*):PARameter:BTESt:NERRors[?]
IAgilentN490xJComplianceMeasurement.NumberOfErrors
MEAS:JTOLerance:COMP(*):PARameter:BTESt:NERRors <NR3>
MEAS:JTOL:COMP(*):PAR:BTESt:NERRors?
This command is needed if the verification method is set to ABSolute. It sets the
number of errored bits. As soon as this number is exceeded, the measurement
proceeds to the next frequency. The setting takes effect when the mode ABSolute
is enabled by means of the MEAS:JTOL:COMP(*):PAR:BTESt command.
The query returns the present number.
MEAS:JTOL:COMP(*):PARameter:STANdard[:SELect][?]
IAgilentN490xJComplianceMeasurement.Standard
MEAS:JTOLerance:COMP(*):PARameter:STANdard[:SELect] <standard enum>
MEAS:JTOLerance:COMP(*):PARameter:STANdard[:SELect]?
This command selects the standard specified by <standard enum> for the Jitter
Tolerance Compliance measurement.
The Serial BERT is equipped with a set of standards provided by Agilent. Every
standard is uniquely identified by an abbreviation (<standard enum>). The
5 SCPI Command Reference
370 Agilent J-BERT N4903B High-Performance Serial BERT
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description