Technical data

MEAS:JTOL:COMP(*):PARameter:FLISt[?]
IAgilentN490xJComplianceMeasurement.GetFrequencyList
IAgilentN490xJComplianceMeasurement.SetFrequencyList
MEAS:JTOLerance:COMP(*):PARameter:FLISt (<NR3>, <NR3>, ...)
MEAS:JTOL:COMP(*):PAR:FLIS?
The command specifies a list of jitter frequencies for the measurement, for example
2.5e4, 5.1e6, 4.755e5, and so on. The unit is Hz. The values must be separated by
commas.
The query returns the present setting, a comma-separated list.
MEAS:JTOL:COMP(*):PARameter:FLISt:FILL
IAgilentN490xJComplianceMeasurement.GenerateFrequencyList()
MEAS:JTOLerance:COMP(*):PARameter:FLISt:FILL <NR3>, <NR3>, <NR1>
This command specifies the start frequency, the stop frequency, and the number
of steps to be used for a Jitter Tolerance Compliance measurement. The frequency
unit is Hz.
The individual frequencies are automatically calculated in logarithmically
equidistant steps. They are returned by the MEAS:JTOL:COMP(*):PAR:FLIS? query.
MEAS:JTOL:COMP(*):PARameter:FLISt:STANdard
IAgilentN490xJComplianceMeasurement.PresetFrequencyList()
MEAS:JTOLerance:COMP(*):PARameter:FLISt:STANdard
Creates a list of test frequencies for a Jitter Tolerance Compliance measurement.
The list contains test frequencies that are recommended for the selected standard:
The start frequency of the standard, the edge frequencies, and the stop frequency.
The minimum number of test frequencies depends on the selected standard. If the
number of steps specified with the MEAS:JTOL:COMP(*):PAR:FLISt:FILL command
is larger than the minimum number of steps, additional frequencies are
interpolated.
5 SCPI Command Reference
368 Agilent J-BERT N4903B High-Performance Serial BERT
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description