Technical data
MEAS:JTOL:CHAR(*):PARameter:BTESt:CLEVel[?]
IAgilentN490xJCharacterizationMeasurement.TargetBERConfidence
MEAS:JTOLerance:CHAR(*):PARameter:BTESt:CLEVel <NR3>
MEAS:JTOL:CHAR(*):PAR:BTESt:CLEV?
This command is needed if the verification method is set to CLEVel. It sets the
confidence level, for example 0.97. The setting takes effect when the confidence
level CLEVel is enabled by means of the MEAS:JTOL:CHAR(*):PAR:BTESt
command.
The query returns the present setting.
MEAS:JTOL:CHAR(*):PARameter:BTESt:NBITs[?]
IAgilentN490xJCharacterizationMeasurement.NumberOfBits
MEAS:JTOLerance:CHAR(*):PARameter:BTESt:NBITs <NR3>
MEAS:JTOL:CHAR(*):PAR:BTESt:NBITs?
This command is needed if the verification method is set to ABSolute. It sets the
number of bits to be captured and compared, for example 5e9. As soon as this
number is exceeded, the measurement proceeds to the next amplitude. The setting
takes effect when the mode ABSolute is enabled by means of the
MEAS:JTOL:CHAR(*):PAR:BTESt command.
The query returns the present number of bits.
MEAS:JTOL:CHAR(*):PARameter:BTESt:NERRors[?]
IAgilentN490xJCharacterizationMeasurement.NumberOfErrors
MEAS:JTOLerance:CHAR(*):PARameter:BTESt:NERRors <NR3>
MEAS:JTOL:CHAR(*):PAR:BTESt:NERRors?
This command is needed if the verification method is set to ABSolute. It sets the
number of errored bits. As soon as this number is exceeded, the measurement
proceeds to the next amplitude. The setting takes effect when the mode ABSolute
is enabled by means of the MEAS:JTOL:CHAR(*):PAR:BTESt command.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 357
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description