Technical data

STATus:QUEStionable:PTRansition[?]
IAgilentN490xStatus.Register (not IVI-compliant)
STATus:QUEStionable:PTRansition
STATus:QUEStionable:PTRansition?
This command sets the transition filter state in the Questionable Status Register.
When this mask is set to "1", positive transitions (logic 0 changing to logic 1) are
allowed to pass. This is the default setting of the instrument. The query returns
the weighted value of the bits that are set to pass positive transitions in the
transition filter. See “Questionable Status Register ” on page 31 for the layout of
the Questionable Status register.
STATus:PROTection Subnode
Data Out, Clock Out, Trigger Out and Aux Out have the same amplifier type. These
amplifiers have a protection circuit that protects the amplifiers against electrical
damage. If a protection circuit is activated the corresponding output will be
disabled.
This subnode has the following SCPI structure:
This subnode has the following commands:
Table 94
Name Description under
:CONDition? “STATus:Protection:CONDition? ” on
page 326
:ENABle[?] “STATus:PROTection:ENABle[?] ” on
page 326
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 325
IVI-COM Equivalent
Syntax
Description