Technical data

Table 3
Bit Mnemonic Description
0 ERR DET Clock loss condition at the
error detector.
1 PAT GEN Clock loss condition at the
pattern generator.
2-15 Not used
Symbol Mode Register
The Symbol Mode Register group indicates whether the error detector has
experienced a 10B symbol alignment loss or 10B symbol alignment done. The
output of this register sets bit 12 (Symbol Mode) in the Questionable Status
Register.
Table 4
Bit Mnemonic Description
0 SYMB ALIGN LOSS 10B symbol alignment loss
condition at the error
detector. The Symb Lock
indicator turns red.
1 SYMB ALIGN DONE 10B symbol alignment
done condition at the error
detector. The Symb Lock
indicator turns green.
2-15 Not used
Questionable Status Register
The Questionable Status Register indicates that a currently running process is of
questionable quality. The output of this register sets bit 3 of the Status Byte.
Recommended Programming Techniques 3
Agilent J-BERT N4903B High-Performance Serial BERT 31