Technical data

Table 70
Name Description under
:DRRatio “[P]FETCh[:SENSe[1]]:DRRatio
Subnode” on page 311
[P]FETCh[:SENSe[1]]:GATE:ELAPsed?
IAgilentN490xEDAccumulation.FetchElapsed (IVI-compliant)
[P]FETCh[:SENSe[1]]:GATE:ELAPsed?
This query returns information about the degree to which the accumulation period
has progressed.
If SENSe[1]:GATE:MANNer TIME is selected, then this command returns the
elapsed time in the accumulation period in units of seconds.
If SENSe[1]:GATE:MANNer ERRors is selected, this command returns the elapsed
errors into the accumulation period.
If SENSe[1]:GATE:MANNer BITS is selected, then this command returns the
elapsed clock bits into the accumulation period.
[P]FETCh[:SENSe[1]]:LOSS:POWer?
[P]FETCh[:SENSe[1]]:LOSS:POWer?
This query returns the total number of seconds that power was lost since the start
of the accumulation period.
This command is not supported.
[P]FETCh[:SENSe[1]]:LOSS:SYNChronizat?
IAgilentN490xEDMeasurement.ReadSecondsSyncLoss (IVI-compliant)
[P]FETCh[:SENSe[1]]:LOSS:SYNChronizat?
This query returns of the total number of seconds for which the incoming pattern
was not synchronized to the reference pattern during the accumulation period.
5 SCPI Command Reference
286 Agilent J-BERT N4903B High-Performance Serial BERT
IVI-COM Equivalent
Syntax
Description
Syntax
Description
IVI-COM Equivalent
Syntax
Description