Technical data

Table 69
Name Description under
Commands
SENSe2:BCOunt? “[P]FETCh:SENSe2:BCOunt? ” on page
282
SENSe2:FREQ[:CW|:FIXed]? “[P]FETCh:SENSe2:FREQuency
[:CW|:FIXed]? ” on page 282
Subnode
[SENSe[1]] “[P]FETCh[:SENSe[1]] Subnode” on
page 283
[P]FETCh:SENSe2:BCOunt?
IAgilentN490xEDMeasurement.ReadBitCount (IVI-compliant)
[P]FETCh:SENSe2:BCOunt?
This query returns the accumulated bit count since the start of the accumulation
period.
[P]FETCh:SENSe2:FREQuency[:CW|:FIXed]?
IAgilentN490xEDMeasurement.ReadClockFrequency (IVI-compliant)
[P]FETCh:SENSe2:FREQuency[:CW|:FIXed]?
This query returns the current frequency of the signal on the clock input. This
measurement is independent of the accumulation period.
5 SCPI Command Reference
282 Agilent J-BERT N4903B High-Performance Serial BERT
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description