Technical data
SENSe2:FREQuency:CDR:TDENsity:MEASure?
SENSe2:FREQuency:CDR:TDENsity:MEASure?
Tries to measure the transition density of the incoming data, and returns the
measured value, in percent. This measurement only succeeds if the CDR is locked;
otherwise, an error is placed in the error queue. The programmer needs to write
the measured value in the hardware via SENS2:FREQ:CDR:TDEN, if desired (this is
not done by the measurement).
SENSe2:FREQuency:CDR:THReshold:AUTO[?]
IAgilentN490xEDCDR.ThresholdAutoTracking (not IVI-compliant)
SENSe2:FREQuency:CDR:THReshold:AUTO 0|1|OFF|ON
SENSe2:FREQuency:CDR:THReshold:AUTO?
Enables/disables the automatic tracking for the CDR threshold. The query returns
the current setting. Default is ON.
The automatic tracking should be disabled for burst applications.
SENSe2:FREQuency:CDR:THReshold[:VALue][?]
IAgilentN490xEDCDR.Threshold (not IVI-compliant)
SENSe2:FREQuency:CDR:THReshold[:VALue] <NR3>
SENSe2:FREQuency:CDR:THReshold[:VALue]?
Sets/gets the manual threshold for CDR bit transitions.
SENSe2:FREQuency:CDR:THReshold:MEASure[?]
IAgilentN490xEDCDR.MeasureAndSetThreshold (not IVI-compliant)
SENSe2:FREQuency:CDR:THReshold:MEASure ONCE
SENSe2:FREQuency:CDR:THReshold:MEASure?