Technical data
SENSe[1]:SFError:ASYMbol:MODe?
For example;
SENS1:SFE:ASYM:MOD REAL
This command configures the settings that control the alignment of the incoming
symbols. The following settings are available:
• SENSe[1]:SFError:ASYMbol:MODe AUT enables automatic alignment.
• SENSe[1]:SFError:ASYMbol:MODe MAN disables automatic alignment and
initiates a re-alignment attempt.
• SENSe[1]:SFError:ASYMbol:MODe REAL re-aligns the symbol alignment, in
case, if it is lost.
This query returns the currently selected mode of the symbol alignment.
SENSe[1]:SFError:CBER:MODe[?]
SENSe[1]:SFError:CBER:MODe SER | FER | FSR | DER | ISR
SENSe[1]:SFError:CBER:MODe?
For example;
SENS1:SFE:CBER:MOD FER
This command sets the calculated BER configuration mode.
This query returns the currently selected mode.
SENSe[1]:SFError:RESult[?]
SENSe[1]:SFError:RESult CBER | SER | FER | FSR | DER | ISR
SENSe[1]:SFError:RESult?
For example;
SENS1:SFE:RES FER
This command configures the results as CBER, SER, FER, FSR, DER or ISR and
displays it in the Results Window.
This query returns the currently selected result.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 269
Description
Syntax
Description
Syntax
Description