Technical data
SENSe[1]:PATTern:UPATtern<n>:LABel[?]
IAgilentN490xEDPatternfile.Description (IVI-compliant)
SENSe[1]:PATTern:UPATtern<n>:LABel <string>
SENSe[1]:PATTern:UPATtern<n>:LABel?
This command sets a description for a user pattern file. The query returns the
description. See “Working with User Patterns in SCPI” on page 63 for information
on using this command.
SENSe[1]:PATTern:UPATtern<n>:USE[?]
IAgilentN490xEDPatternfile.Alternate (IVI-compliant)
SENSe[1]:PATTern:UPATtern<n>:USE STRaight | APATtern
SENSe[1]:PATTern:UPATtern<n>:USE?
STR | APAT
This command defines whether a user pattern file should be a straight pattern or
an alternate pattern:
• STRaight
The pattern is repeatedly output.
• APATtern
The pattern is composed of two halves. The output depends on various other
commands; see “How the Serial BERT Uses Alternate Patterns” on page 55 for
more information.
The default is set to have a length of 128 bits for each half pattern; all bits are set
to zero and the trigger is set to occur on the A/B changeover. See “Working with
User Patterns in SCPI” on page 63 for information on using this command.
5 SCPI Command Reference
258 Agilent J-BERT N4903B High-Performance Serial BERT
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Return Range
Description