Technical data

The query returns the time to which the gate period is set.
SENSe[1]:GATE[:STATe][?]
IAgilentN490xEDAccumulation.GetState
IAgilentN490xEDAccumulation.Start IAgilentN490xEDAccumulation.Abort
SENSe[1]:GATE[:STATe] 0 | 1 | OFF | ON
SENSe[1]:GATE[:STATe]?
This command turns accumulation on or off.
N O T E
Previous commands that have altered the configuration of the instrument might not
have settled. In order to ensure that the GATE ON command is not executed until
conditions have settled, it is strongly recommended that the frequency be allowed
to stabilize prior to the GATE ON command, and then be followed by a
synchronization search. For more information, see Allowing Serial BERT to Settle
- Concepts” on page 22.
N O T E
When GATE:MODE SINGle is executed, the GATE[:STATe]ON command is an
overlapped command. For more information, see “Overlapped and Sequential
Commands ” on page 80 the .
The query returns the current state of the accumulation gating.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 247
IVI-COM Equivalent
Syntax