Technical data
SENSe[1]:GATE:PERiod:BITS[?]
IAgilentN490xEDAccumulation.Bits (IVI-compliant)
SENSe[1]:GATE:PERiod:BITS <NR3>
SENSe[1]:GATE:PERiod:BITS?
When GATE:MANNer is set to BITS, the duration of the accumulation period is set
in clock bits (or periods). <NR3> can be a value between 1E7 and 1E15 in decade
steps.
Executing this command invalidates all past results.
The query returns the number of bits to which the gate period is set.
SENSe[1]:GATE:PERiod:ERRors[?]
IAgilentN490xEDAccumulation.Errors (IVI-compliant)
SENSe[1]:GATE:PERiod:ERRors <NR3>
SENSe[1]:GATE:PERiod:ERRors?
When GATE:MANNer is set to ERRors, the command sets the duration of the
accumulation period in bit errors. Values of 10, 100 and 1000 are permitted.
Executing this command invalidates all past results.
The query returns the number of errors to which the gate period is set.
SENSe[1]:GATE:PERiod[:TIMe][?]
IAgilentN490xEDAccumulation.Time (IVI-compliant)
SENSe[1]:GATE:PERiod[:TIME] <NR3>
SENSe[1]:GATE:PERiod[:TIME]?
When GATE:MANNer is set to TIME, the duration of the accumulation period is
set in seconds. Neither a value less than 1 second nor greater than 8639999
seconds (99 days, 23 hours, 59 minutes and 59 seconds) is permitted.
The command causes all past results to be labelled as invalid.
5 SCPI Command Reference
246 Agilent J-BERT N4903B High-Performance Serial BERT
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description