Technical data
SENSe[1]:GATE:MANNer <Manner>
SENSe[1]:GATE:MANNer?
TIME | ERR | BITS
This command sets the manner in which the accumulation period is controlled.
<Manner> can be one of the following:
• TIME
The error detector performs SINGle and REPetitive accumulation periods that
are controlled by elapsed time.
• ERRors
The error detector performs SINGle and REPetitive accumulation periods that
are controlled by the accumulation of bit errors.
• BITS
The error detector performs SINGle and REPetitive accumulation periods that
are controlled by the accumulation of clock bits.
The query returns the current manner of accumulation.
SENSe[1]:GATE:MODE[?]
IAgilentN490xEDAccumulation.ActivationMode (IVI-compliant)
SENSe[1]:GATE:MODE <Mode>
SENSe[1]:GATE:MODE?
MAN | SING | REP
The command sets the accumulation period mode <Mode>, which can be either:
• MANual
The accumulation is started manually (by sending GATE:STATe ON).
• SINGle
Errors are accumulated over one accumulation period.
• REPetitive
The accumulation loops, according to the setting of :GATe:MANNer.
Executing this command invalidates all past results. The query returns the current
accumulation mode.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 245
Syntax
Return Range
Description
IVI-COM Equivalent
Syntax
Return Range
Description