Technical data

SENSe[1]:GATE:MANNer <Manner>
SENSe[1]:GATE:MANNer?
TIME | ERR | BITS
This command sets the manner in which the accumulation period is controlled.
<Manner> can be one of the following:
TIME
The error detector performs SINGle and REPetitive accumulation periods that
are controlled by elapsed time.
ERRors
The error detector performs SINGle and REPetitive accumulation periods that
are controlled by the accumulation of bit errors.
BITS
The error detector performs SINGle and REPetitive accumulation periods that
are controlled by the accumulation of clock bits.
The query returns the current manner of accumulation.
SENSe[1]:GATE:MODE[?]
IAgilentN490xEDAccumulation.ActivationMode (IVI-compliant)
SENSe[1]:GATE:MODE <Mode>
SENSe[1]:GATE:MODE?
MAN | SING | REP
The command sets the accumulation period mode <Mode>, which can be either:
MANual
The accumulation is started manually (by sending GATE:STATe ON).
SINGle
Errors are accumulated over one accumulation period.
REPetitive
The accumulation loops, according to the setting of :GATe:MANNer.
Executing this command invalidates all past results. The query returns the current
accumulation mode.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 245
Syntax
Return Range
Description
IVI-COM Equivalent
Syntax
Return Range
Description