Technical data

SENSe[1]:ELOCation:BEADdress[?]
IAgilentN490xEDErrorLocation.BitAddress (not IVI-compliant)
SENSe[1]:ELOCation:BEADdress <NR1>
SENSe[1]:ELOCation:BEADdress?
This command only has effect if the BER location mode is set up to locate errors
on a single bit in the pattern. The command sets the position of the bit that is to
be monitored for errors.
The query returns the current value.
See the Serial BERT User's Guide and “SENSe[1]:BLOCk[?]” on page 233.
If the query is used with Error Location Capture, the return value depends on the
Error Location Capture status:
If an error has been captured, the bit position of the errored bit is returned.
If no error has been captured since the last start of error location capture, the
last set BEADdress is returned.
See the Serial BERT User's Guide and “Using Error Location Capture - Procedures”
on page 46 for additional information.
SENSe[1]:ELOCation:VERBose?
IAgilentN490xEDErrorLocation.ReadState (not IVI-compliant)
SENSe[1]:ELOCation:VERBose?
This query returns the current state of the error location capture measurement.
The following responses are possible:
ELOC__NOT_STARTED_YET
Status: Stopped; ELOC not started since last instrument power-up.
ELOC__RUNNING
Status: Running; ELOC is running, searching for an errored bit.
ELOC__ERROR_DETECTED
Status: Running; ELOC has detected an errored bit. The data is being evaluated.
ELOC__SUCCESS
5 SCPI Command Reference
236 Agilent J-BERT N4903B High-Performance Serial BERT
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description