Technical data

SENSe[1]:BLOCk[?]
IAgilentN490xEDErrorLocation.Mode (not IVI-compliant)
SENSe[1]:BLOCk ON | OFF | 0 | 1 | BEADdress | WPATtern | BLOCk
SENSe[1]:BLOCk?
This command configures the error location feature of the instrument. It is only
available for user-defined patterns (selected with PATTern:SELect UPATtern or
PATTern:SELect FILename). Only the errors within the defined location are counted
by the instrument.
The command has the following options:
WPATtern
When this is selected, all errors that occur through the entire pattern are
counted. OFF and 0 (NULL) have the same effect.
BEADdress
Only errors that occur at the specified bit address are counted. Use :ELOC:BEAD
to specify the bit address.
BLOCk
Errors that occur within the specified bit address range are counted. The bit
address range is defined with the :BLOC:BST and BLOC:BLEN commands. ON
and 1 have the same effect.
The query returns the following:
0 (equivalent to WPATtern)
1 (equivalent to BLOCk)
BEAD
SENSe[1]:BLOCk:BSTart[?]
IAgilentN490xEDErrorLocation.BlockStart (not IVI-compliant)
SENSe[1]:BLOCk:BSTart <NR1>
SENSe[1]:BLOCk:BSTart?
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 233
IVI-COM Equivalent
Syntax
Description
Parameters
Return Values
IVI-COM Equivalent
Syntax