Technical data
Table 50
Name Description under
:PATTern “SENSe[1]:PATTern Subnode” on page
248
:VOLTage “SENSe[1]:VOLTage Subnode” on page
264
:SFError “SENSe[1]:SFError Subnode” on page
266
SENSe[1]:LOGGing[?]
IAgilentN490xEDLogging.Enabled (IVI-compliant)
SENSe[1]:LOGGing 0 | 1 | OFF | ON
SENSe[1]:LOGGing?
This command allows you to save accumulated results in a file for later analysis.
The query form returns whether or not the accumulated results will be saved in a
log file.
N O T E
This command is not precisely the same as 716xxB command. The ONCE parameter
is no longer supported.
SENSe[1]:LOGGing:FILename[?]
IAgilentN490xEDLogging.Filename (IVI-compliant)
SENSe[1]:LOGGing:FILename <Filename>
SENSe[1]:LOGGing:FILename?
This command specifies the file to which logged data is sent. The query returns
the filename to which the logged data is sent.
You have to explicitly enable logging with SENSe[1]:LOGGing.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 227
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description