Technical data

This subsystem has the following commands:
Table 49
Name Description under
:COUPling[?] “INPut[1]:COUPling[?] ” on page 221
:DELay[?] “INPut[1]:DELay[?] ” on page 221
:POLarity[?] “INPut[1]:POLarity[?] ” on page 222
:TERMination[?] “INPut[1]:TERMination[?] ” on page
222
:STATe[?] “INPut[1]:STATe[?] ” on page 223
:CMODe[?] “INPut[1]:CMODe[?] ” on page 223
INPut[1]:COUPling[?]
IAgilentN490xEDDataIn.TerminationEnabled (IVI-compliant)
INPut[1]:COUPling AC | DC
INPut[1]:COUPling?
The command enables or disables the source of the termination voltage:
DC: Enables the termination voltage
AC: Disables the termination voltage
The query returns the current state.
N O T E
Non-differential operation of the error detector's Data In port requires a termination
voltage.
INPut[1]:DELay[?]
IAgilentN490xEDDataIn.Delay (IVI-compliant)
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 221
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent