Technical data
:SOUR8:SPEC:LEV(*)?
A suffix attached to the SCPI ..:LEVel(*) command is used to address two jitter
levels. The SCPI command ...:LEVel1 <NR3> changes the level of the internal RJ
resource and and the command …:LEVel2 <NR3> changes the level of the internal
BUJ resource. Both levels are specified in rms.
SOURce8[:JITTer]:SPECtrally:FILTer[:LPASs][:STATe][?]
IAgilentN490xJSPECtrally.LowPassFilterEnable
SOURce8:SPECtrally:FILTer ON | OFF | 1 | 0
SOUR8:SPEC:FILT?
Turns the 500 MHz low-pass filter for spectrally random jitter on or off.
The query returns the present setting (0 | 1).
SOURce8[:JITTer]:RSSClocking Subnode
The residual Spread Spectrum Clocking (rSSC) setting influences the pattern
generator's outputs. When rSSC is enabled, it can have an impact on the Data Out,
Clock Out, Trigger Out and Aux Out of the generator.
The instrument has one internal source for Spread Spectrum Clocking (SSC),
residual Spread Spectrum Clock (rSSC) and Sinusoidal Jitter (SJ). So, residual
Spread Spectrum Clocking is not available if the Spread Spectrum Clocking or the
Sinusoidal jitter is enabled.
This subnode has the following SCPI structure:
This subnode has the following commands:
5 SCPI Command Reference
218 Agilent J-BERT N4903B High-Performance Serial BERT
Description
IVI-COM Equivalent
Syntax
Description