Technical data

SOUR8[:JITT]:PER2:SWE:STAN[:SEL]?
The command defines the predefined number of standards that define a periodic
sweep curve. These periodic jitter curves corresponds to the compliance curves
also used by the Jitter Tolerance Compliance measurement.
The query returns the present setting.
SOURce8[:JITTer]:SINusoidal Subnode
The Sinusoidal Jitter setting influences the pattern generator's outputs. When SJ
is enabled, it can have an impact on the Data Out, Clock Out, Trigger Out and Aux
Out of the generator.
The instrument has one internal source for Spread Spectrum Clocking (SSC) and
residual Spread Spectrum Clocking (rSSC). So, Sinusoidal Jitter is not available if
residual Spread Spectrum Clocking is enabled.
This subnode has the following SCPI structure:
This subnode has the following commands:
Table 46
Name Description under
[:STATe][?] “SOURce8[:JITTer]:SINusoidal[:STATe]
[?] ” on page 215
:CLOCk[?] “SOURce8[:JITTer]:SINusoidal:CLOCk
[?]” on page 215
:FREQuency[?] “SOURce8
[:JITTer]:SINusoidal:FREQuency[?] ” on
page 215
5 SCPI Command Reference
214 Agilent J-BERT N4903B High-Performance Serial BERT
Description