Technical data
SOUR8:RAND:FILT:SEL?
Selcts between the 500MHz (LP500 default) low-pass filter and the 100MHz
(LP100) low-pass filter.
The query returns the present setting (LP500 | LP100).
SOURce8[:JITTer]:RANDom:FILTer:HPASs[:STATe][?]
IAgilentN490xJRandom.HighPassFilterEnable
SOURce8:RANDom:FILTer:HPAS ON | OFF | 1 | 0
SOUR8:RAND:FILT:HPAS?
Turns the 10 MHz high-pass filter for random jitter on or off.
The query returns the present setting (0 | 1).
SOURce8[:JITTer]:RANDom:LEVel[?]
IAgilentN490xJRandom.Amplitude, GetMaxAmplitude(), GetMinAmplitude()
SOURce8:RANDom:LEVel <NR3>
SOUR8:RAND:LEVel? MIN | MAX
The command sets the root mean square (rms) RJ level in unit intervals (UI).
The query returns the present setting or the applicable min/max values in UI.
SOURce8[:JITTer]:RANDom:CFACtor?
SOURce8:RANDom:CFACtor?
This query returns the crest factor of the generated random jitter distribution. The
present implementation of random jitter generates a fixed crest factor of 14.
The crest factor is defined as the peak-to-peak value divided by the rms value.
5 SCPI Command Reference
202 Agilent J-BERT N4903B High-Performance Serial BERT
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
Syntax
Description