Technical data
OUTPut[5]:EIDLe[:STATe]
IAgilent:EIDLe[:STATe]
OUTPut[5]::EIDLe[:STATe] 0 | 1 | ON | OFF
OUTPut[5]::EIDLe[:STATe]?
This command enables/disables the electrical idle feature Aux-Output. The Error
Input can be used to provide an electrical idle (or out of band signaling) signal at
the Aux Output. An external generated signal e.g. by a pulse generator at the Error
Input controls the output stream at the Aux Output. If this STATe is set to ON, error
insertion is not possible.
OUTPut[5]:DATA:XOVer[?]
IAgilentN490xPGOutput.Crossover (IVI-compliant)
OUTPut[5]:DATA:XOVer <NR1.>
OUTPut[5]:DATA:XOVer? [MINimum | MAXimum]
The command sets the eye crossover of the pattern generator's Data Out port.
Crossover can only be changed in NRZ signal mode.
The query returns the current crossover setting.
SENSe6 Subsystem
SENSe6 Subsystem - Reference
The SENSe6 Subsystem represents the pattern generator's Clock In ports.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 191
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description