Technical data
SOURce5:VOLTage[:LEVel][:IMMediate]:OFFSet?
The command sets the offset value of the pattern generator's Clock Out port in
Volts. The query returns the offset value of the pattern generator's Clock Out port
in Volts.
SOURce5:VOLTage:LLEVel[?]
IAgilentN490xPGAuxOutVoltage.LogicLevel (not IVI-compliant)
SOURce5:VOLTage:LLEVel <Family>
SOURce5:VOLTage:LLEVel?
<Family>: ECL | LVPECL | SCFL | LVDS | CML | CUSTom
N O T E
Selecting CUSTom has no effect.
ECL | LVPECL | SCFL | LVDS | CML | CUSTom
The command sets the output level appropriate for the specified logic family. The
query returns the currently used logic family.
N O T E
If any of the voltage parameters have been modified, CUSTom will be returned by
the query, even if the parameter has been set back to the default.
SOURce5:VOLTage:PROTection:RESet[?]
IAgilentN490xPGAuxOutVoltage.ProtectionReset (not IVI-compliant)
SOURce5:VOLTage:PROTection:RESet[?]
SOURce5:VOLTage:PROT:RES?
This command reset the protection circuit to normal operation at the Aux Output
channel.
5 SCPI Command Reference
188 Agilent J-BERT N4903B High-Performance Serial BERT
Description
IVI-COM Equivalent
Syntax
Input Parameters
Return Range
Description
IVI-COM Equivalent
Syntax
Description