Technical data
SOURce3:TRIGger:VOLTage[:LEVel][:IMMediate]:LOW <NR3>
SOURce3:TRIGger:VOLTage[:LEVel][:IMMediate]:LOW?
The command sets the DC low output level of the pattern generator's Clock Out
port in Volts. The query returns the DC low output level of the pattern generator's
Clock Out port in Volts.
SOURce3:TRIGger:VOLTage[:LEVel][:IMMediate]:OFFSet[?]
IAgilentN490xPGTrigVoltage.VOffset (IVI-compliant)
SOURce3:TRIGger:VOLTage[:LEVel][:IMMediate]:OFFSet <NR3>
SOURce3:TRIGger:VOLTage[:LEVel][:IMMediate]:OFFSet?
The command sets the offset value of the pattern generator's Clock Out port in
Volts. The query returns the offset value of the pattern generator's Clock Out port
in Volts.
SOURce3:TRIGger:VOLTage:LLEVel[?]
IAgilentN490xPGTrigger.LogicLevel (not IVI-compliant)
SOURce3:TRIGger:VOLTage:LLEVel <Family>
SOURce3:TRIGger:VOLTage:LLEVel?
<Family>: ECL | LVPECL | SCFL | LVDS | CML | CUSTom
N O T E
Selecting CUSTom has no effect.
ECL | LVPECL | SCFL | LVDS | CML | CUSTom
The command sets the output level appropriate for the specified logic family. The
query returns the currently used logic family.
N O T E
If any of the voltage parameters have been modified, CUSTom will be returned by
the query, even if the parameter has been set back to the default.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 181
Syntax
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Input Parameters
Return Range
Description