Technical data

Table 37
Name Description under
:TRIGger:VOLTage:LLEVel[?] “SOURce3:TRIGger:VOLTage:LLEVel[?]
” on page 181
:TRIGger:VOLTage:PROTection:RESet[?] “SOURce3:VOLTage:PROTection:RESet
[?] ” on page 182
:TRIGger:VOLTage:PROTection[:STATe]
[?]
“SOURce3:VOLTage:PROTection
[:STATe][?] ” on page 182
:TRIGger:OUTPut:COUPling[?] “SOURce3:TRIGger:OUTPut:COUPling[?]
” on page 182
:TRIGger:OUTPut:TERMination[?] “SOURce3:TRIGger:OUTPut:TERMinatio
n[?] ” on page 182
N O T E
See “How the Serial BERT Sends Triggers” on page 58 for details about trigger
signals are generated.
SOURce3:TRIGger[:MODE][?]
IAgilentN490xPGTrigger.Mode (IVI-compliant)
SOURce3:TRIGger[:MODE] DCLock | PATTern | SEQuence
SOURce3:TRIGger[:MODE]?
DCL | PATT | SEQ
This command sets the pattern generator's Trigger/Ref Clock Out to one of three
modes:
DCLock: Sets the Trigger/Ref Clock Out to divided clock mode. The divider is set
with the command SOURce3:TRIGger:DCDRatio.
PATTern: Sets the Trigger/Ref Clock Out to pattern trigger mode. The subsequent
commands apply.
SEQuence: Sets the Trigger/Ref Clock Out to sequence trigger mode. If a user-
defined sequence is used, only DCL or SEQ can be chosen.
5 SCPI Command Reference
176 Agilent J-BERT N4903B High-Performance Serial BERT
IVI-COM Equivalent
Syntax
Return Range
Description
Parameters