Technical data
OUTPut[1]:CLOCk:TERMination?
This command is applicable for N4916B. The CLK-IN input is AC coupled and thus
does not provide a threshold voltage, but it does have an adjustable termination
voltage. The available range is from -2V to +2V. When setting the :INPut:COUPling
to DC, then the internal termination voltage is applied to the input signal.
This query returns the present termination voltage.
N O T E
This command is only valid if the coupling is set to DC (see “OUTPut
[1]:CLOCk:COUPling[?]” on page 160).
OUTPut[1]:CLOCk:COUPling[?]
OUTPut[1]:CLOCk:COUPling <NR3>
OUTPut[1]:CLOCk:COUPling?
This command is applicable for N4916B. The command enables or disables the
source of the termination voltage:
• DC: Enables the termination voltage
• AC: Disables the termination voltage
The query returns the present coupling state.
SOURce9 Subsystem
SOURce9 Subsystem - Reference
The SOURce9 Subsystem represents the pattern generator's Clock Out port. It is
also used to control the optional Spread Spectrum Clocking function (SSC).
5 SCPI Command Reference
160 Agilent J-BERT N4903B High-Performance Serial BERT
Description
Syntax
Description