Technical data
This command is applicable for M8061A. The command sets the termination
voltage of the electrical idle. The termination voltage parameter can only be change
if the output is off (disabled).
The query returns the present setting.
OUTPut[1]:MULTiplexer:TCONfig[?]
OUTPut[1]:MULTiplexer:TCONfig BALance | UNBalance
OUTPut[1]:MULTiplexer:TCONfig?
This command is applicable for M8061A. The command is used to enable/disable
termination voltage or in other words to select the termination model for output
frontend. The termination model can be set to:
• BALanced – balanced or differential output
• UNBalanced – unbalanced with respect to ground (termination voltage)
The termination model parameter can only be changed if the output of data out is
off (disabled).
The query returns the present setting.
OUTPut[1]:MULTiplexer:TVOLtage[?]
OUTPut[1]:MULTiplexer:TVOLtage <NRf>
OUTPut[1]:MULTiplexer:TVOLtage?
This command is applicable for M8061A. The command sets the external
termination voltage at the data output. The termination voltage parameter can only
be changed if the output is off (disabled).
The query returns the present setting.
5 SCPI Command Reference
156 Agilent J-BERT N4903B High-Performance Serial BERT
Description
Syntax
Description
Syntax
Description