Technical data

is selected, the amplitude range changes will be done automatically by changing
the amplitude value.
The query returns the present state.
OUTPut[1]:MULTiplexer:VOLTage:PROTection[:STATe]?
OUTPut[1]:MULTiplexer:VOLTage:PROTection[:STATe]?
This query is applicable for M8061A. The query returns the state of the protection
circuit at the Data Out channel of M8061A. The "ON" state indicates a triggered
protection circuit or in other words the output is disabled and no signal comes out.
OUTPut[1]:MULTiplexer:EIDLe[:STATe][?]
OUTPut[1]:MULTiplexer:EIDLe OFF | IDLE | EXTernal
OUTPut[1]:MULTiplexer:EIDLe?
This command is applicable for M8061A. The command sets the state of electrical
idle to either OFF, IDLE or External.
The query returns the present state.
OUTPut[1]:MULTiplexer:EIDLe:THReshold[?]
OUTPut[1]:MULTiplexer:THReshold <NRf>
OUTPut[1]:MULTiplexer:THReshold?
This command is applicable for M8061A. The command sets the threshold level of
the electrical idle.
The query returns the present setting.
OUTPut[1]:MULTiplexer:EIDLe:TVOLtage[?]
OUTPut[1]:MULTiplexer:EIDLe:TVOLtage <NRf>
OUTPut[1]:MULTiplexer:EIDLe:TVOLtage?
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 155
Syntax
Description
Syntax
Description
Syntax
Description
Syntax