Technical data
OUTPut[1]:MULTiplexer:VOLTage:HIGH[?]
OUTPut[1]:MULTiplexer:VOLTage:HIGH <NRf>
OUTPut[1]:MULTiplexer:VOLTage:HIGH?
This command is applicable for M8061A. The command sets the high output level
in volts.
The query returns the low output level in volts.
OUTPut[1]:MULTiplexer:VOLTage:LOW[?]
OUTPut[1]:MULTiplexer:VOLTage:LOW <NRf>
OUTPut[1]:MULTiplexer:VOLTage:LOW?
This command is applicable for M8061A. The command sets the low output level
in volts.
The query returns the low output level in volts.
OUTPut[1]:MULTiplexer:VOLTage:RANGe[:SELect][?]
OUTPut[1]:MULTiplexer:VOLTage:RANGe:SELect 1
OUTPut[1]:MULTiplexer:VOLTage:RANGe:SELect?
This command is applicable for M8061A. The command allows the user to select
the set of predefined amplitude range from the provided list. This feature will be
enabled only if "Auto Range" option is disabled.
The query returns the present state.
OUTPut[1]:MULTiplexer:VOLTage:RANGe:AUTO[?]
OUTPut[1]:MULTiplexer:VOLTage:RANGe:AUTO ON | OFF
OUTPut[1]:MULTiplexer:VOLTage:RANGe:AUTO?
This command is applicable for M8061A. The command enables/disables
automatic tracking of the amplitude ranges. If "Auto Range" is not selected,
amplitude value can only be varied within in the specified range. If "Auto Range"
5 SCPI Command Reference
154 Agilent J-BERT N4903B High-Performance Serial BERT
Syntax
Description
Syntax
Description
Syntax
Description
Syntax
Description