Technical data
Sets the duty cycle of a repetitive pulse waveform (like in RZ or R1 signal modes).
Duty cycle value <NR1> in % of the clock period. Valid range is 0 ... 100, default is
50.
The query returns the current setting.
OUTPut[1]:DELay[?]
IAgilentN490xPGOutput.Delay (IVI-compliant)
OUTPut[1]:DELay <NR3>
OUTPut[1]:DELay?
This command sets the delay of the active edge of the clock output relative to the
pattern generator's Data Out port. The units are seconds. The response returns
the current data to clock delay value.
This command has restrictions for frequencies under 620 Mbits/s. See for details.
See the User Guide (or online Help) for details.
OUTPut[1]:FORMat[?]
IAgilentBertPGOutput.SignalMode
OUTPut[1]:FORMat <NRZ | RZ | R1>
OUTPut[1]:FORMat?
Sets the pulse format ("Signal Mode") to NRZ, RZ, or R1.
The query returns the current setting as a string.
N O T E
If you wish to use the RZ or R1 formats in conjunction with a directly supplied
external clock (not a reference clock), you must specify the external clock as
"manual" (instead of "automatic"). For details see “SENSe6:MODE ” on page
193.
In RZ or R1 mode, the crossover cannot be changed.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 137
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description