Technical data

[SOURce[1]]:VOLTage[:LEVel]:LLEVel <Family>
[SOURce[1]]:VOLTage[:LEVel]:LLEVel?
<Family>: ECL | LVPECL | SCFL | LVDS | CML | CUSTom
ECL | LVPECL | SCFL | LVDS | CML | CUST
N O T E
Selecting CUSTom has no effect.
The command sets the output level appropriate for the specified logic family. The
query returns the currently used logic family.
N O T E
If any of the voltage parameters have been modified, CUSTom will be returned by
the query, even if the parameter has been set back to the default.
[SOURce[1]]:VOLTage:PROTection:RESet
IAgilentN490xPGOutVoltage.ProtectionReset (not IVI-compliant)
[SOURce[1]]:VOLTage:PROTection:RESet
[SOURce[1]]:VOLTage:PROT?
This command reset the protection circuit to normal operation at the Data Output
channel.
[SOURce[1]]:VOLTage:PROTection[:STATe][?]
IAgilentN490xPGOutVoltage.ProtectionState (not IVI-compliant)
[SOURce[1]]:VOLTage:PROTection:[:STATe]?
This query gets the state of the protection circuit at the Data Output channel. The
retun values are 'OFF' or 'ON'. The 'ON' state indicates a triggered protection
circuit or in other words the output is disabled and no signal comes out.
5 SCPI Command Reference
132 Agilent J-BERT N4903B High-Performance Serial BERT
Syntax
Input Parameters
Return Range
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description