Technical data
N O T E
For the UPATtern<n> commands, <n> can be in the range 0 - 12. 0 (zero) is used
to select the current pattern, 1 - 12 selects one of the user patterns in the memory.
[SOURce[1]]:PATTern:UPATtern<n>[:LENGth][?]
IAgilentN490xPGPatternfile.Length (IVI-compliant)
[SOURce[1]]:PATTern:UPATtern<n>[:LENGth] <NR1>
[SOURce[1]]:PATTern:UPATtern<n>[:LENGth]?
This command sets the length of the selected user pattern. The query returns the
length of the user pattern. If an alternate pattern is selected (:USE APATtern), the
LENGth command sets the length of each half of the pattern.
Note that the :DATA command automatically sets the length of the pattern.
See “Working with User Patterns in SCPI” on page 63 for information on using this
command.
[SOURce[1]]:PATTern:UPATtern<n>:LABel[?]
IAgilentN490xPGPatternfile.Description (IVI-compliant)
[SOURce[1]]:PATTern:UPATtern<n>:LABel <string>
[SOURce[1]]:PATTern:UPATtern<n>:LABel?
The command sets the description of the pattern. The query returns the description
of the pattern.
See “Working with User Patterns in SCPI” on page 63 for information on using this
command.
[SOURce[1]]:PATTern:UPATtern<n>:USE[?]
IAgilentN490xPGPatternfile.Alternate (IVI-compliant)
[SOURce[1]]:PATTern:UPATtern<n>:USE STRaight | APATtern
[SOURce[1]]:PATTern:UPATtern<n>:USE?
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 125
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax