Technical data
[SOURce[1]]:PATTern:SEQuence:RCL
IAgilentN490xPGSequence.Load()
SOURce[1]:PATTern:SEQuence:RCL <FileIdentifier>
This command recalls (loads) a sequence that has been stored in a file by means
of the SOURce[1]:PATTern:SEQuence:SAVE command.
The FileIdentifier must include path and file name (for example,
"C:\<Instrument
Model>\Sequences\seq01.seq"
).
N O T E
The contents of the specified file overwrites the present SequenceExpression. See
also “SequenceExpression for User-Defined Sequences” on page 112. You can
download the new SequenceExpression to the pattern generator with the command
SOURce[1]:PATTern:SELect SEQuence.
Recall may fail if the SequenceExpression references a pattern file that is not
available.
SOURce[1]:PATTern:SEQuence:SAVe
IAgilentN490xPGSequence.Save()
SOURce[1]:PATTern:SEQuence:SAVe <FileIdentifier>
This command saves a SequenceExpression in a file.
The FileIdentifier must include path and file name (for example,
C:\<Instrument
Model>\Sequences\seq01.seq
).
N O T E
The standard extension of a sequence file is
.seq
. The file is ASCII-coded and
contains the SequenceExpression. See also “SequenceExpression for User-Defined
Sequences” on page 112.
[SOURce[1]]:PATTern:SEQuence:STATe?
IAgilentN490xPGSequence.State()
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 117
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent