Technical data
• AuxInHi | AuxInLo: Sequence starts when Auxiliary Input is high or low.
• AuxInRising | AuxInFalling: Sequence starts with a rising or falling edge at the
Auxiliary Input.
• Manual: Sequence starts when the [SOURce[1]]:PATTern:SEQ:EVENt ONCE
command is received.
Example:
Start= AuxInRising
Defines the contents of a block and the On/Off of the associated sequence trigger.
Syntax:
Block #= <PatternType1.0> [, TrigOn |, TrigOff]
Block #= <PatternType1.1> [, TrigOn |, TrigOff]
The range of "#" is 1 to 120.
N O T E
A sequence consists of up to 120 blocks that can be looped. The sum of blocks and
counted loops must not exceed 120.
Input parameters:
• For version 1.0, the PatternType is one of the following:
None| PRBS#[, <Length>] | <UserFilename>[,A|,B] | P0[, <Length>] | P1,
[<Length>] | CL/#[, <Length>]
For version 1.1, the PatternType is one of the following:
None,X| PRBS#,Y,[, <Length>] | <UserFilename>[,A|,B],Y, [, <Length>] |
P0,Y[, <Length>]| P1, Y[, <Length>] | CL/#,Y[, <Length>]
Where:
X = None
Y = PRBS#| <UserFilename>[,A|,B] | P0| P1,| CL/#
N O T E
Length is related to master channel (Data Out) while the other channel (Aux Data)
will always follow this length.
N O T E
'None' is only possible for both channel synchronically.
Example: Block 1= None, None, 311040, TrigOn is possible
Example: Block 1=None, P1, 311040, TrigOn is not possible
The default length of a block is 512 bits. Therefore, this parameter can be
omitted.
5 SCPI Command Reference
114 Agilent J-BERT N4903B High-Performance Serial BERT
Block #=