Technical data

[SOURce[1]]:PATTern[:SELect] <Source>
[SOURce[1]]:PATTern[:SELect]?
<Source>: PRBS<n> | PRBN<n> | ZSUBstitut<n> | MDENsity<n> |
UPATtern<n> | FILename, <string> | SEQuence | PRBS23P
PRBS<n> | PRBN<n> | ZSUB<n> | MDEN<n> | UPAT | SEQ | PRBS23P
This command defines the type of pattern being generated. The parameter is
retained for backwards compatibility and may be one of the following:
PRBS<n> <n> = 7, 10, 11, 15, 23, 31
PRBN<n> <n> = 7, 10,11,13, 15, 23
ZSUBstitut<n> <n> = 7, 10,11,13, 15, 23
UPATtern<n> <n> = 1 through 12
MDENsity<n> <n> = 7, 10,11,13, 15, 23
FILename, <string>
SEQuence
PRBS23P
ZSUBstitut: Zero SUBstitution; used for defining PRBN patterns in which a block
of bits is replaced by a block of zeros. The length of the block is defined by
“[SOURce[1]]:PATTern:ZSUBstitut[:ZRUN][?]” on page 106.
MDENsity: Mark DENsity; used for defining a PRBN pattern in which the user can
set the mark density. The mark density is set with “[SOURce
[1]]:PATTern:MDENsity[:DENSity][?] ” on page 104.
UPATtern<n>: User PATtern; used to define the contents of a pattern store. For
the Serial BERT, <n> can be 1 to12.
FILename: A parameter that allows the remote user to load a user pattern from
the instrument's disk drive. This is the preferred mechanism for loading user
patterns in the Serial BERT.
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 105
Syntax
Input Parameters
Return Range
Description