Technical data

96 Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
6 Status Reporting
The Status Byte Register, the Standard Event Status Register group, and
the Output Queue are defined as the Standard Status Data Structure
Model in IEEE 488.2- 1987. IEEE 488.2 defines data structures, commands,
and common bit definitions for status reporting. There are also
oscilloscope- defined structures and bits.
The status reporting structure consists of the registers shown here.
Table 4 lists the bit definitions for each bit in the status reporting data
structure.
Figure 2 Status Reporting Overview Block Diagram
Table 4 Status Reporting Bit Definition
Bit Description Definition
PON Power On Indicates power is turned on.
URQ User Request Not Used. Permanently set to zero.
CME Command Error Indicates if the parser detected an error.
EXE Execution Error Indicates if a parameter was out of range or
was inconsistent with the current settings.