:WAVeform:YORigin?
Query :WAVeform:YORigin?
The :WAVeform:YORigin? query returns the y- origin voltage value for the
currently specified source. The voltage value returned is the voltage value
represented by the waveform data digital code 00000.
For BYTE and WORD data, and voltage waveforms, it is the voltage at
digital code zero.
For ASCii data format, the YORigin is the Y-axis value at the center of
the data range. Data range is returned in the Y increment.
Returned Format
[:WAVeform:YORigin] <real_value><NL>
<real_value> A real number in exponential format.
Example This example places the current Y- origin value in the numeric variable,
varCenter, then prints the contents of the variable to the computer's
screen.
myScope.WriteString ":SYSTEM:HEADER OFF" ' Response headers off.
myScope.WriteString ":WAVEFORM:YORIGIN?"
varCenter = myScope.ReadNumber
Debug.Print FormatNumber(varCenter, 0)
See Also For more information on BYTE and WORD formats, see "Understanding
WORD and BYTE Formats" on page 883.
You can obtain the Y- origin value through the :WAVeform:PREamble?
query.
NOTE
A "Waveform data is not valid" error occurs when there is no data available for a channel.
When this occurs, a zero value is returned.