Technical data

894 Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
32 Waveform Commands
See Also :WAVeform:DATA?
Y Origin The Y origin is the Y-axis value at level zero. For voltage waveforms, it is
the voltage at level zero. (See the :WAVeform:YORigin? query.)
Y Reference The Y reference is the level associated with the Y origin. It is at this level
that the Y origin is defined. In this oscilloscope, this value is always zero.
(See the :WAVeform:YREFerence? query.)
Coupling The input coupling of the waveform. The coupling value is ignored when
sent to the oscilloscope in the preamble. (See the :WAVeform:COUPling?
query.)
X Display Range The X display range is the X-axis duration of the waveform that is
displayed. For time domain waveforms, it is the duration of time across
the display. (See the :WAVeform:XRANge? query.)
X Display Origin The X display origin is the X-axis value at the left edge of the display. For
time domain waveforms, it is the time at the start of the display. This
value is treated as a double precision 64-bit floating point number. (See
the :WAVeform:XDISplay? query.)
Y Display Range The Y display range is the Y-axis duration of the waveform which is
displayed. For voltage waveforms, it is the amount of voltage across the
display. (See the :WAVeform:YRANge? query.)
Y Display Origin The Y-display origin is the Y-axis value at the center of the display. For
voltage waveforms, it is the voltage at the center of the display. (See the
:WAVeform:YDISplay? query.)
Date The date that the waveform was acquired or created.
Time The time that the waveform was acquired or created.
Frame Model # The model number of the frame that acquired or created this waveform.
The frame model number is ignored when it is sent to an oscilloscope in
the preamble.
Acquisition Mode The acquisition sampling mode of the waveform. (See :ACQuire:MODE.)
Complete The complete value is the percent of time buckets that are complete. The
complete value is ignored when it is sent to the oscilloscope in the
preamble. (See the :WAVeform:COMPlete? query.)
X Units The X-axis units of the waveform. (See the :WAVeform:XUNits? query.)
Y Units The Y-axis units of the waveform. (See the :WAVeform:YUNits? query.)
Band Pass The band pass consists of two values that are estimates of the
maximum and minimum bandwidth limits of the source waveform. The
bandwidth limit is computed as a function of the selected coupling and
filter mode. (See the :WAVeform:BANDpass? query.)
Table 27 Waveform Preamble Elements (continued)
Element Description