<Y increment> The Y increment is the duration between Y- axis levels. For voltage
waveforms, it is the voltage corresponding to one level. If the value is zero
then no data has been acquired. (See the :WAVeform:YINCrement? query.)
<Y origin> The Y origin is the Y- axis value at level zero. For voltage waveforms, it is
the voltage at level zero. If the value is zero then no data has been
acquired. (See the :WAVeform:YORigin? query.)
<Y reference> The Y reference is the level associated with the Y origin. It is at this level
that the Y origin is defined. In this oscilloscope, this value is always zero.
(See the :WAVeform:YREFerence? query.)
<coupling> 0 for AC coupling. 1 for DC coupling. 2 for DCFIFTY coupling. 3 for
LFREJECT coupling.
<X display
range>
The X display range is the X-axis duration of the waveform that is
displayed. For time domain waveforms, it is the duration of time across
the display. If the value is zero then no data has been acquired. (See the
:WAVeform:XRANge? query.)
<X display
origin>
The X display origin is the X-axis value at the left edge of the display. For
time domain waveforms, it is the time at the start of the display. This
value is treated as a double precision 64- bit floating point number. If the
value is zero then no data has been acquired. (See the
:WAVeform:XDISplay? query.)
<Y display
range>
The Y display range is the Y- axis duration of the waveform which is
displayed. For voltage waveforms, it is the amount of voltage across the
display. If the value is zero then no data has been acquired. (See the
:WAVeform:YRANge? query.)
<Y display
origin>
The Y- display origin is the Y- axis value at the center of the display. For
voltage waveforms, it is the voltage at the center of the display. If the
value is zero then no data has been acquired. (See the
:WAVeform:YDISplay? query.)
<date> A string containing the date in the format DD MMM YYYY, where DD is
the day, 1 to 31; MMM is the month; and YYYY is the year.
<time> A string containing the time in the format HH:MM:SS:TT, where HH is the
hour, 0 to 23, MM is the minutes, 0 to 59, SS is the seconds, 0 to 59, and
TT is the hundreds of seconds, 0 to 99.
<frame_
model_#>
A string containing the model number and serial number of the
oscilloscope in the format of MODEL#:SERIAL#.
<acquisition
_mode>
0 for RTIMe mode. 1 for ETIMe mode. 2 not used. 3 for PDETect.
<completion> The completion value is the percent of time buckets that are complete. The
completion value is ignored when it is sent to the oscilloscope in the